STEREOTACTIC BODY RADIATION THERAPY – DOSIMETRY AND MECHANICAL PREPARATION OF LINEAR ACCELERATOR

M. Szymański, M. Bojarojć, A. Jabłońska, B. Kuśmierska-Piątek

Pages: 59-62

DOI: 10.21175/RadProc.2023.11

The aim of this study is to emphasize the importance of proper dosimetric and mechanical preparation of the linear accelerator and to highlight the significance of this step in ensuring the efficacy and safety of stereotactic body radiation therapy (SBRT). Performance verification of two Varian VitalBeam linear accelerators was executed in order to monitor the output consistency, MLC repeatability and radiation isocenter location and size. Dose output consistency measurements were conducted using phantoms and dosimetric equipment from PTW. The analysis of MLC positioning tests, isocenter size and location were performed by in-house scripts based on pylinac tools. These tests were conducted on a weekly basis. Evaluation of the results of tests performed over a 90-week period showed a positive trend in output consistency. However, no discernible trend or pattern was observed in the other tests. In conclusion, the authors suggest that performing dosimetric and mechanical checks on the linear accelerator before conducting SBRT treatment is an effective way to maintain a safe treatment environment.
  1. H. Pan, D.R. Simpson, L.K. Mell, A.J. Mundt, J.D. Lawson, “A survey of stereotactic body radiotherapy use in the United States”, Cancer, 117(19), 4566-72, 2011. http://doi.org/10.1002/cncr.26067
  2. A. Dimitriadis, “Assessing the dosimetric and geometric accuracy of stereotactic radiosurgery”, Department of Physics Faculty of Engineering and Physical Science University of Surrey, December 2016.
  3. J. Malicki, K. Ślosarek, “Planowanie Leczenia i Dozymetria w Radioterapii (Tom 2)”, Via Medica, 2018, pp. 889-907.
  4. Rozporządzenie Ministra Zdrowia z dnia 12 grudnia 2022 w sprawie testów eksploatacyjnych urządzeń radiologicznych i urządzeń pomocniczych. Retrieved from: https://isap.sejm.gov.pl/isap.nsf/DocDetails.xsp?id= WDU20220002759
  5. Obwieszczenie Ministra Zdrowia z dnia 22 grudnia 2014 r. w sprawie ogłoszenia wykazu wzorcowych procedur radiologicznych z zakresu radioterapii onkologicznej. Retrieved from: https://sip.lex.pl/akty-prawne/dzienniki- resortowe/ogloszenie-wykazu-wzorcowych-procedur- radiologicznych-z-zakresu-34930484
  6. IAEA HUMAN HEALTH REPORTS No. 18. National Networks for Radiotherapy Dosimetry Audits. Structure, Methodology, Scientific Procedures. INTERNATIONAL ATOMIC ENERGY AGENCY VIENNA, 2023. Retrieved from: https://www-pub.iaea.org/MTCD/publications/PDF/PUB1964_web.pdf
  7. Technical specifications of radiotherapy equipment for cancer treatment. ISBN 978-92-4-001998-0, World Health Organization, 2021. Retrieved from: https://apps.who.int/iris/bitstream/handle/10665/33 9912/9789240019980-eng.pdf
  8. P. Kukołowicz, W. Ślusarczyk-Kacprzyk, P. Wesołowska, M. Szymański, A. Walewska, I. Grabska, “Rola audytu dozymetrycznego w bezpiecznej radioterapii / The role of dosimetric audit in safe radiotherapy”, Inżynier i Fizyk Medyczny, 12,291-294, 2023.
  9. M. Szymański, M. Piziorska, O. Madetko, W. Ślusarczyk-Kacprzyk, W. Bulski, “Dosimetry audit of the CyberKnife accelerator with the SHANE phantom”, Polish Journal of Medical Physics and Engineering, The Journal of Polish Society of Medical Physics, 27(3), 207-212, 2021. http://doi.org/10.2478/pjmpe-2021-0025
  10. Q. Fan, S. Zhou, Y. Lei, S. Li, M. Zhang, “A Quality Assurance Approach for Linear Accelerator Mechanical Isocenters with Portal Images”, Int. J. Medical Physics, Clinical Engineering and Radiation Oncology, 7(1), 100-114, 2018. http://doi.org/10.4236/ijmpcero.2018.71009
  11. Pylinac Documentation, Version 3.11. Retrieved from: https://pylinac.readthedocs.io/en/latest/
  12. International Atomic Energy Agency. IAEA Safety Standards Series No. 398: Quality Assurance for Radiotherapy. Vienna: IAEA, 2018.
  13. M.A Bolt, C.H. Clark, T. Chen, A. Nisbet, “A multi- centre analysis of radiotherapy beam output measurement”, Physics and Imaging in Radiation Oncology, 4, 39-43, 2017. http://doi.org/10.1016/j.phro.2017.12.001
  14. H. Szweda, K. Graczyk, D. Radomiak, K. Matuszewski, “Comparison of three different phantoms used for Winston-Lutz test with Artiscan software”, Reports of Practical Oncology and Radiotherapy, 25(3), 351-354, 2020. http://doi.org/10.1016/j.rpor.2020.03.003
M. Szymański, M. Bojarojć, A. Jabłońska, B. Kuśmierska-Piątek, "Stereotactic body radiation therapy – dosimetry and mechanical preparation of linear accelerator", RAD Conf. Proc., vol. 7, 2023, pp. 59-62; http://doi.org/10.21175/RadProc.2023.11