STEREOTACTIC BODY RADIATION THERAPY – DOSIMETRY AND MECHANICAL PREPARATION OF LINEAR ACCELERATOR
M. Szymański, M. Bojarojć, A. Jabłońska, B. Kuśmierska-Piątek
Pages: 59-62
DOI: 10.21175/RadProc.2023.11
Abstract |
References | Cite This | Full Text (PDF)
The aim of this study is to emphasize the importance of proper dosimetric and mechanical preparation of
the linear accelerator and to highlight the significance of this step in ensuring the efficacy and safety of stereotactic
body radiation therapy (SBRT). Performance verification of two Varian VitalBeam linear accelerators was executed
in order to monitor the output consistency, MLC repeatability and radiation isocenter location and size. Dose output
consistency measurements were conducted using phantoms and dosimetric equipment from PTW. The analysis of
MLC positioning tests, isocenter size and location were performed by in-house scripts based on pylinac tools. These
tests were conducted on a weekly basis. Evaluation of the results of tests performed over a 90-week period showed a
positive trend in output consistency. However, no discernible trend or pattern was observed in the other tests. In
conclusion, the authors suggest that performing dosimetric and mechanical checks on the linear accelerator before
conducting SBRT treatment is an effective way to maintain a safe treatment environment.
- H. Pan, D.R. Simpson, L.K. Mell, A.J. Mundt, J.D.
Lawson, “A survey of stereotactic body radiotherapy
use in the United States”, Cancer, 117(19), 4566-72,
2011.
http://doi.org/10.1002/cncr.26067
- A. Dimitriadis, “Assessing the dosimetric and geometric
accuracy of stereotactic radiosurgery”, Department of
Physics Faculty of Engineering and Physical Science
University of Surrey, December 2016.
- J. Malicki, K. Ślosarek, “Planowanie Leczenia i
Dozymetria w Radioterapii (Tom 2)”, Via Medica,
2018, pp. 889-907.
- Rozporządzenie Ministra Zdrowia z dnia 12 grudnia
2022 w sprawie testów eksploatacyjnych urządzeń
radiologicznych i urządzeń pomocniczych.
Retrieved from:
https://isap.sejm.gov.pl/isap.nsf/DocDetails.xsp?id=
WDU20220002759
- Obwieszczenie Ministra Zdrowia z dnia 22 grudnia 2014
r. w sprawie ogłoszenia wykazu wzorcowych procedur
radiologicznych z zakresu radioterapii onkologicznej.
Retrieved from:
https://sip.lex.pl/akty-prawne/dzienniki-
resortowe/ogloszenie-wykazu-wzorcowych-procedur-
radiologicznych-z-zakresu-34930484
- IAEA HUMAN HEALTH REPORTS No. 18. National
Networks for Radiotherapy Dosimetry Audits.
Structure, Methodology, Scientific Procedures.
INTERNATIONAL ATOMIC ENERGY AGENCY
VIENNA, 2023.
Retrieved from:
https://www-pub.iaea.org/MTCD/publications/PDF/PUB1964_web.pdf
- Technical specifications of radiotherapy equipment for
cancer treatment. ISBN 978-92-4-001998-0, World
Health Organization, 2021.
Retrieved from:
https://apps.who.int/iris/bitstream/handle/10665/33
9912/9789240019980-eng.pdf
- P. Kukołowicz, W. Ślusarczyk-Kacprzyk, P. Wesołowska,
M. Szymański, A. Walewska, I. Grabska, “Rola audytu
dozymetrycznego w bezpiecznej radioterapii / The role
of dosimetric audit in safe radiotherapy”, Inżynier i
Fizyk Medyczny, 12,291-294, 2023.
- M. Szymański, M. Piziorska, O. Madetko, W.
Ślusarczyk-Kacprzyk, W. Bulski, “Dosimetry audit of
the CyberKnife accelerator with the SHANE phantom”,
Polish Journal of Medical Physics and Engineering,
The Journal of Polish Society of Medical Physics, 27(3),
207-212, 2021.
http://doi.org/10.2478/pjmpe-2021-0025
- Q. Fan, S. Zhou, Y. Lei, S. Li, M. Zhang, “A Quality
Assurance Approach for Linear Accelerator Mechanical
Isocenters with Portal Images”, Int. J. Medical Physics,
Clinical Engineering and Radiation Oncology, 7(1),
100-114, 2018.
http://doi.org/10.4236/ijmpcero.2018.71009
- Pylinac Documentation, Version 3.11.
Retrieved from:
https://pylinac.readthedocs.io/en/latest/
- International Atomic Energy Agency. IAEA Safety
Standards Series No. 398: Quality Assurance for
Radiotherapy. Vienna: IAEA, 2018.
- M.A Bolt, C.H. Clark, T. Chen, A. Nisbet, “A multi-
centre analysis of radiotherapy beam output
measurement”, Physics and Imaging in Radiation
Oncology, 4, 39-43, 2017.
http://doi.org/10.1016/j.phro.2017.12.001
- H. Szweda, K. Graczyk, D. Radomiak, K. Matuszewski,
“Comparison of three different phantoms used for
Winston-Lutz test with Artiscan software”, Reports of
Practical Oncology and Radiotherapy, 25(3), 351-354,
2020.
http://doi.org/10.1016/j.rpor.2020.03.003
M. Szymański, M. Bojarojć, A. Jabłońska, B. Kuśmierska-Piątek, "Stereotactic body radiation therapy – dosimetry and mechanical preparation of linear accelerator", RAD Conf. Proc., vol. 7, 2023, pp. 59-62; http://doi.org/10.21175/RadProc.2023.11
|